Dental HandPiece Maintainance Remove Water blockage / clog , Dental Airotar Water Spray Problems Fixed By Nexus Medodent (NMD) #dental #handpiecemaintainance
NMD want to help you get the best performance and longevity from your handpieces. In order to so we have compiled this list, which if followed will help.
Dental handpieces are delicate instruments and fail prematurely if not properly maintained. The majority of handpiece failures are due to poor maintenance & care, because of limited knowledge in this area.
The following tips will help you get longer service from your NMD handpieces.
1. External Cleaning
Clean the outer casing of handpiece with a soft to medium bristle brush and alcohol or warm water.
Do Not submerge handpiece in water or any type of chemical.
Do Not place in ultrasonic cleaner.
2. Thermal Washer Disinfector
If using a washer disinfector, when washer cycle is complete always remove handpiece, never leave handpiece in a washer e.g. overnight.
3. Lubricate & Clean Internally
Using a good quality service oil & the correct adaptor, insert into handpiece and spray for approx 2 seconds until oil comes from the head. You cannot over oil your handpieces.
The spindle/chuck of handpieces should be lubricated once a day with the pointed lubricant adaptor.
4. Expel Debris and Excess Lubricant
Insert bur and connect handpiece to tubing. Run handpieces for 10-20 seconds after oiling and before sterilizing to expel debris and excess lubricant. This will also help keep the handpieces clean internally.
5. Sterilize
Remove bur from handpiece, wipe down exterior & insert into pouch/bag. Sterilize handpiece as per manufacturers instructions. Always allow handpiece to go through the dry cycle on your sterilizer (if there is a dry cycle on it). Remove handpiece from autoclave when process is completed. Never leave handpiece in autoclave.
6. Let Cool & Dry
Upon completion of the sterilization process remove handpiece from autoclave as soon as possible. Store with head upright until cool and dry internally. Never use a handpiece before it has cooled down and always allow it to cool naturally. Do not cool handpieces under a running tap or in cold water.
7. Check Air Supply
Check the air pressure going to your handpieces. The pressure should not exceed 2.8bar/40psi. If the pressure is too high this will cause premature bearing failure. Check that the air supply is clean and dry.
Have your service engineer check your compressor and air/water filters in the dental unit on a regular basis.
Always Clean, Lubricate & sterilize your handpieces after every patient.
Following these tips, correct care and maintenance will avoid premature handpiece failure. We also wrote a guide to help choosing the right dental handpiece, if you’re interested!
Spray nozzles and water filter are cleaned carefully with the nozzle cleaner to remove dirt and deposits
Save money and down time by performing minor handpiece repair problems in office. Clearing dental handpiece water line and spray cover obstructions in office may be easier than you think. With guitar string / file and a small utility syringe, most handpiece water line problems can be solved
In many handpieces, the most tortuous path water takes is in the vicinity of the handpiece head, so you may only need to insert the wire a quarter to half an inch to achieve the desired result. In addition, many times the culprit is amalgam or powder which enters at the spray ports when peddle is released, so only the water exit point is clogged.
Depending on your brand of handpiece, and the number of spray ports it handpiece repairemploys, it’s important to understand the flow path water takes prior to ending up as atomized spray. On a single port handpiece, it’s likely that there is one solid tube from start to finish, and many times water clean out wire can be pushed through the entire length, resulting in a good clearing of all obstructions. On multi port spray models, such as the those manufactured by different brands and others, water enters via a single line, but then mixes with air in a ring header prior to exiting the handpiece head. In these cases, it is impossible for cleanout wire to travel the entire length of the water line. However, many times it is just the exit ports which are clogged, and the above process can be used to avoid a trip to the handpiece repair shop