Building Fluke 5700A calibrator - Part 1
xDevs_com
0:00 / 0:00
Building Fluke 5700A calibrator - Part 1
4 912 просмотров · Трансляция закончилась 7 лет назад
xDevs_com
3,57 тыс. подписчиков
4 912 просмотров · Трансляция закончилась 7 лет назад
Live-stream for Fluke 5700A Rev2 project, repairing random calibrator modules and parts.
07:00 - Calibration demo for Fluke 87V DMM with Fluke 5720A calibrator
12:00 - Mainboards overview for second Fluke 5700A build project
18:00 - Look on old abused 5700A front panel
26:00 - Overview on 5700A service manual book
27:30 - First look on broken A3 mainboard board
40:40 - Desoldering mains voltage switches to gain access to PCB
1:03:00 - Cutting PCB damaged part up to 4 layers deep
1:07:40 - Delaminated PCB pieces removed
1:11:16 - Cutting more stuff. Beware the audio noise.
1:21:30 - Cleaning mess
1:32:30 - Restoring broken traces with jump-wires
2:10:00 - Solder switches back to board
2:30:00 - Removing transformer frame from chassis
2:32:30 - Talk about Agilent 4142B parametric analyzer
2:38:15 - Fluke mains filter
2:45:00 - Fluke voodoo transformers
3:02:00 - First smoke test, first mains fuse is dead :)
3:15:00 - Checking transformer winding resistances
3:45:00 - Checking board resistances
3:49:20 - Close look on original fuse
3:57:30 - Stealing parts from good calibrator for test
4:01:00 - Look on A20 and A19 boards
4:05:50 - Second fuse is dead
4:13:50 - Power on with HV connector disconnected, first messages on display
TiN is blowing up fuses, milling PCB and doing live troubleshooting of Fluke 5700A inguard A3 outguard motherboard 6-layer PCBA.
Worklog available here: xdevs.com/fix/hulk2/