Don't Fall Into This Output Impedance Trap!

Keysight Labs

0:00 / 0:00

Don't Fall Into This Output Impedance Trap!

44 466 просмотров · 8 л. назад
Keysight Labs
120 тыс. подписчиков
44 466 просмотров · 8 л. назад
Function Gen says 1V, Scope says 2V. Which one's right? Click to subscribe! ► http://bit.ly/Scopes_Sub ◄ The function generator is set to have a 1 V output, but the oscilloscope measures a 2 Volt signal. Why? When set to 50 ohm mode, the function generator actually outputs double the specified input voltage. It does this because it expects to see a 50 ohm load on the output, creating a 50/50 voltage divider circuit. Questions or comments? Ask me on Twitter: @DanielBogdanoff   / danielbogdanoff   Learn more about using oscilloscopes: http://oscilloscopelearningcenter.com Check out the EEs Talk Tech podcast:    • EEs Talk Tech - An Electrical Engineering ...   The digital storage oscilloscope we used: http://www.keysight.com/en/pcx-x20520... #oscilloscope #oscilloscopes #electronics #electricalengineering #impedance #impedancematching #50ohm #50ohmimpedance #highimpedance #outputimpedance #functiongenerator