This is the third video in a four-part series on masking basics.
Initial masking level (AC):
Air conduction threshold of the non-test ear + 10 dB safety pad
Steps:
1. Find your initial masking level
2. Increase the stimulus level until you get a response
3. Raise the masker until you get no response or a 15-20 dB plateau
4. Repeat 2-3 until you find the plateau
Series videos:
1: Knowing when to mask: Air conduction
2: Knowing when to mask: Bone conduction
3. How to mask: Air conduction
4. How to mask: Bone conduction
In this video we're going over how to mask when measuring air conduction hearing thresholds.